NCV FIELD ATLASEvidence-based nerve conduction learning Open in the interactive 3D atlas
Motor study Lower limb

Fibular (peroneal) motor nerve conduction study

Place G1 over the center of EDB and stimulate sequentially at the ankle, below the fibular head and above the fibular head.

01 · Placement

G1, G2, stimulation and ground landmarks

G1 active recording electrode

G1: center of the EDB

Place G1 over the center of the extensor digitorum brevis.

Table 2 · p.373 ↗
G2 reference electrode

G2: distal to the fifth MTP joint

Place G2 distal to the fifth metatarsophalangeal joint.

Table 2 · p.373 ↗
Stimulation site

Ankle stimulation: lateral to the tibialis anterior tendon

Stimulate at the ankle lateral to the tibialis anterior tendon, 8 cm from G1.

Table 2 · p.373 ↗
Stimulation site

Below fibular head: posteroinferior

Place the below-head stimulation site posteroinferior to the fibular head.

Table 2 · p.373 ↗
Stimulation site

Above fibular head: 10 cm proximal

Place the above-head stimulation site 10 cm proximal to the below-head site and just medial to the biceps femoris tendon.

Table 2 · p.373 ↗
Ground region

Ground electrode region

AANEM places the ground between the stimulating and recording electrodes. The model shows a usable anterolateral calf region, not a single mandatory point.

Table 2 · p.373 ↗

02 · Reference values

Adult normal limits

MeasureReference limitBasis
Distal CMAP amplitude ≥1.3 mV 3rd percentile lower limit
CV ankle–below head ≥38 m/s 3rd percentile lower limit
CV across fibular head ≥42 m/s 3rd percentile lower limit
Distal motor latency ≤6.5 ms 97th percentile upper limit
Amplitude drop across head ≤25% 97th percentile upper limit
CV drop across head ≤6 m/s or 12% 97th percentile upper limit

03 · Technique

Acquisition conditions

Fixed distance G1 to ankle: 8 cm; across-fibular-head segment: 10 cm
Sensitivity 5 mV/div
Sweep 5 ms/div
Filter 2–3 Hz–10 kHz
Temperature Maintain dorsal foot skin temperature >31°C
Ground Place the ground between the stimulating and recording electrodes

Primary evidence

Method-matched AANEM reference data

Adult fibular motor reference values recorded from EDB; see the source table for age- and height-stratified limits.

Open Table 4, p.375 ↗