Motor study Lower limb

Fibular (peroneal) motor nerve conduction study

Place G1 over the center of EDB and stimulate sequentially at the ankle, below the fibular head and above the fibular head.

01 · Placement

G1, G2, stimulation and ground landmarks

G1 active recording electrode

G1: center of the EDB

Place G1 over the center of the EDB muscle belly on the dorsolateral foot, adjusting to obtain an initial negative CMAP deflection.

Table 2 · p.373 ↗
G2 reference electrode

G2: distal to the fifth MTP joint

Place G2 just distal to the fifth MTP joint over the distal EDB tendon, completely off the muscle belly.

Table 2 · p.373 ↗
Stimulation site

Ankle stimulation: lateral to the tibialis anterior tendon

Stimulate at the anterior ankle 8 cm proximal to G1, immediately lateral to the tibialis anterior tendon; orient the cathode distally.

Distance · 8 cm from G1

Table 2 · p.373 ↗
Stimulation site

Below fibular head: posteroinferior

Stimulate the common fibular nerve posteroinferior to the fibular head where it approaches the fibular neck.

Distance · Variable—measure the S1–S2 segment on the patient

Table 2 · p.373 ↗
Stimulation site

Above fibular head: 10 cm proximal

Stimulate about 10 cm proximal to S2, just medial to the biceps femoris tendon at the lateral popliteal fossa.

Distance · 10 cm proximal to S2

Table 2 · p.373 ↗
Optional accessory fibular study

Optional accessory fibular stimulation: behind lateral malleolus

For the optional accessory fibular study, stimulate just posterior to the lateral malleolus while continuing to record from EDB; a reproducible CMAP supports an accessory branch.

Distance · Variable—stimulate just posterior to the lateral malleolus

Accessory fibular nerve technique · p.25 ↗
Ground region

Ground electrode region

Place the ground on the anterolateral lower leg between the ankle stimulation site and the proximal fibular-head sites.

Table 2 · p.373 ↗

02 · Reference values

Adult normal limits

MeasureReference limitBasis
Distal CMAP amplitude ≥1.3 mV 3rd percentile lower limit
CV ankle–below head ≥38 m/s 3rd percentile lower limit
CV across fibular head ≥42 m/s 3rd percentile lower limit
Distal motor latency ≤6.5 ms 97th percentile upper limit
Amplitude drop across head ≤25% 97th percentile upper limit
CV drop across head ≤6 m/s or 12% 97th percentile upper limit

03 · Technique

Acquisition conditions

Fixed distance G1 to ankle: 8 cm; across-fibular-head segment: 10 cm; accessory site is anatomical
Sensitivity 5 mV/div
Sweep 5 ms/div
Limb position S4 is optional when an accessory fibular response is suspected
Filter 2–3 Hz–10 kHz
Temperature Maintain dorsal foot skin temperature >31°C
Ground Place the ground between the stimulating and recording electrodes

Primary evidence

Method-matched AANEM reference data

Adult fibular motor reference values recorded from EDB; see the source table for age- and height-stratified limits.

Open Table 4, p.375 ↗