G1: EIP muscle belly
With the forearm pronated, place G1 over the extensor indicis proprius belly approximately two fingerbreadths proximal to the ulnar styloid.
Radial motor study · p.18 ↗Record over EIP and stimulate at the forearm, elbow, below the spiral groove and above the spiral groove.
01 · Placement
With the forearm pronated, place G1 over the extensor indicis proprius belly approximately two fingerbreadths proximal to the ulnar styloid.
Radial motor study · p.18 ↗Place G2 over the ulnar styloid on the distal extensor indicis tendon, at least 3 cm distal to G1 and completely off the muscle belly.
Distance · At least 3 cm distal to G1
Radial motor technique · p.19 ↗Stimulate over the dorsal ulna 4–6 cm proximal to G1 with the cathode directed toward the recording electrode.
Distance · 4–6 cm proximal to G1
Radial motor study · p.18 ↗Stimulate the radial nerve at the lateral elbow in the groove between the distal biceps tendon and brachioradialis.
Distance · Variable—measure the S1–S2 segment on the patient
Radial motor study · p.18 ↗Stimulate at the lateral mid-arm between biceps and triceps, immediately distal to the radial nerve's spiral-groove segment.
Distance · Variable—measure the S2–S3 segment on the patient
Radial motor study · p.18 ↗Stimulate over the proximal posterior humerus near the axilla, proximal to the spiral groove; measure this segment directly on the patient.
Distance · Variable—measure the S3–S4 segment on the patient
Radial motor study · p.18 ↗Place the ground on the dorsal forearm between the distal forearm stimulation site and the EIP recording electrodes.
Table 2 · p.373 ↗02 · Reference values
| Measure | Reference limit | Basis |
|---|---|---|
| Distal CMAP amplitude | ≥2 mV | AANEM educational reference |
| Conduction velocity | ≥49 m/s | Method-matched segment |
| Distal motor latency | ≤2.9 ms | 4–6 cm distal distance |
03 · Technique
Primary evidence
Radial motor reference values recorded from extensor indicis with a 4–6 cm distal segment.