G1 active recording electrode
G1: along the line toward the radial artery
Place G1 along the line toward the radial artery.
Table 2 · p.373 ↗Record along the line toward the radial artery and stimulate just lateral to the distal biceps tendon.
01 · Placement
Place G1 along the line toward the radial artery.
Table 2 · p.373 ↗Use a bar electrode with G2 3 cm distal to G1.
Table 2 · p.373 ↗Stimulate just lateral to the distal biceps tendon, 10 cm from G1.
Table 2 · p.373 ↗AANEM places the ground between the stimulating and recording electrodes. The model shows a usable lateral forearm region, not a single mandatory point.
Table 2 · p.373 ↗02 · Reference values
| Measure | Reference limit | Basis |
|---|---|---|
| SNAP onset-to-peak | ≥5 µV | 3rd percentile lower limit |
| SNAP peak-to-peak | ≥6 µV | 3rd percentile lower limit |
| Onset latency | ≤2.5 ms | 97th percentile upper limit |
03 · Technique
Primary evidence
Adult lateral antebrachial cutaneous sensory reference values.