G1 active recording electrode
G1: index finger
Record from digit II with G1; the AANEM reference values use an antidromic technique.
Table 2 · p.373 ↗Place G1 on the index finger; stimulate at the wrist between the flexor carpi radialis and palmaris longus tendons.
01 · Placement
Record from digit II with G1; the AANEM reference values use an antidromic technique.
Table 2 · p.373 ↗Place G2 4 cm distal to G1.
Table 2 · p.373 ↗Stimulate at the wrist between the flexor carpi radialis and palmaris longus tendons, 14 cm from G1.
Table 2 · p.373 ↗Stimulate the palm midway between the 14 cm wrist site and G1, 7 cm from G1.
Table 2 · p.373 ↗AANEM places the ground between the stimulating and recording electrodes. The model shows a usable palmar wrist region, not a single mandatory point.
Table 2 · p.373 ↗02 · Reference values
| Measure | Reference limit | Basis |
|---|---|---|
| Wrist SNAP onset-to-peak | ≥11 µV | 3rd percentile lower limit |
| Wrist SNAP peak-to-peak | ≥13 µV | 3rd percentile lower limit |
| Wrist onset latency | ≤3.3 ms | 97th percentile upper limit |
| Wrist peak latency | ≤4.0 ms | 97th percentile upper limit |
| Palm onset latency | ≤1.6 ms | 7 cm segment |
| Palm peak latency | ≤2.3 ms | 7 cm segment |
| Conduction velocity | ≥43 m/s | Published 3rd percentile |
03 · Technique
Primary evidence
Adult antidromic median sensory reference values for digit II; amplitude is also affected by age and BMI.