Ulnar motor nerve conduction study

Place G1 over the hypothenar eminence and stimulate sequentially at the wrist, below elbow, above elbow and axilla.

01 · Placement

G1, G2, stimulation and ground landmarks

G1 active recording electrode

G1: hypothenar eminence

Place G1 over the hypothenar eminence.

Table 2 · p.373 ↗
G2 reference electrode

G2: just distal to the fifth MCP joint

Place G2 just distal to the fifth metacarpophalangeal joint.

Table 2 · p.373 ↗
Stimulation site

Wrist stimulation: just radial to the FCU tendon

Stimulate at the wrist just radial to the flexor carpi ulnaris tendon, 8 cm from G1.

Table 2 · p.373 ↗
Stimulation site

Below elbow: 4 cm distal to the medial epicondyle

Flex the elbow to 90° and stimulate 4 cm distal to the medial epicondyle.

Table 2 · p.373 ↗
Stimulation site

Above elbow: 10 cm proximal along the curved path

From the below-elbow site, measure 10 cm proximally along the curve behind the medial epicondyle, just volar to the triceps.

Table 2 · p.373 ↗
Stimulation site

Axilla: 10 cm proximal to the above-elbow site

Stimulate in the axilla 10 cm proximal to the above-elbow site.

Table 2 · p.373 ↗
Ground region

Ground electrode region

AANEM places the ground between the stimulating and recording electrodes. The model shows a usable ulnar forearm region, not a single mandatory point.

Table 2 · p.373 ↗

02 · Reference values

Adult normal limits

MeasureReference limitBasis
Distal CMAP amplitude ≥7.9 mV 3rd percentile lower limit
CV below elbow ≥52 m/s 3rd percentile lower limit
CV across elbow ≥43 m/s 3rd percentile lower limit
CV above elbow ≥50 m/s 3rd percentile lower limit
Distal motor latency ≤3.7 ms 97th percentile upper limit
CV drop across elbow ≤15 m/s or 23% 97th percentile upper limit

03 · Technique

Acquisition conditions

Fixed distance G1 to wrist: 8 cm; across-elbow segment: 10 cm
Sensitivity 5 mV/div
Sweep 2 ms/div
Limb position Elbow flexed 90°
Filter 2–3 Hz–10 kHz
Temperature Maintain dorsal hand skin temperature >32°C
Ground Place the ground between the stimulating and recording electrodes

Primary evidence

Method-matched AANEM reference data

Adult ulnar motor reference values, including the across-elbow conduction-velocity drop limit.

Open Table 4, p.375 ↗