Sensory study Upper limb

Ulnar sensory nerve conduction study

Place G1 on the fifth digit and stimulate at the wrist just radial to the flexor carpi ulnaris tendon.

01 · Placement

G1, G2, stimulation and ground landmarks

G1 active recording electrode

G1: fifth digit

Place the active ring electrode around the proximal phalanx of digit V at the proximal fifth MCP region.

Table 2 · p.373 ↗
G2 reference electrode

G2: 4 cm distal to G1

Place the reference ring electrode 4 cm distal to G1 along the little finger, toward the distal MCP and fingertip.

Distance · 4 cm distal to G1

Table 2 · p.373 ↗
Stimulation site

Wrist stimulation: just radial to the FCU tendon

Stimulate at the volar wrist 14 cm proximal to G1, immediately radial to the flexor carpi ulnaris tendon; orient the cathode distally.

Distance · 14 cm from G1

Table 2 · p.373 ↗
Optional proximal sensory stimulation

Optional below-elbow stimulation

For optional proximal sensory conduction, flex the elbow and stimulate the ulnar nerve in the retrocondylar course just below the medial epicondyle.

Distance · Variable—measure the S1–S2 segment on the patient

Ulnar sensory proximal conduction · p.15 ↗
Optional proximal sensory stimulation

Optional above-elbow stimulation

For optional proximal sensory conduction, stimulate above the medial epicondyle along the curved ulnar-nerve path between biceps and triceps.

Distance · Variable—measure the S2–S3 segment on the patient

Ulnar sensory proximal conduction · p.15 ↗
Ground region

Ground electrode region

Place the ground over the ulnar wrist crease between the wrist stimulation site and the digit-V recording electrodes.

Table 2 · p.373 ↗

02 · Reference values

Adult normal limits

MeasureReference limitBasis
SNAP onset-to-peak ≥10 µV 3rd percentile lower limit
SNAP peak-to-peak ≥9 µV 3rd percentile lower limit
Onset latency ≤3.1 ms 97th percentile upper limit
Peak latency ≤4.0 ms 97th percentile upper limit
Conduction velocity ≥45 m/s Published 3rd percentile

03 · Technique

Acquisition conditions

Fixed distance G1 to wrist stimulation: 14 cm; measure optional proximal segments on the patient
Sensitivity 20 µV/div
Limb position Below- and above-elbow sensory stimulation is optional and rarely performed
Filter 20 Hz–2 kHz
Sweep 1 ms/div
Temperature Maintain dorsal hand skin temperature >32°C
Ground Place the ground between the stimulating and recording electrodes

Primary evidence

Method-matched AANEM reference data

Adult antidromic ulnar sensory reference values for digit V; amplitude is also affected by age and BMI.

Open Table 3, p.374 ↗